NTU stands for Nephelometric Turbidity Unit. It is the most widely used measurement for turbidity — the cloudiness or haziness of a liquid caused by suspended particles that scatter light passing through it.
The higher the NTU value, the more particles are present and the more turbid the liquid appears. A reading close to zero means the liquid is virtually clear. A reading in the hundreds or thousands means the liquid is visibly cloudy or opaque.
In filtration, NTU does two jobs. It tells you how much work the filter has ahead of it before you start. And it tells you whether the filter is doing that work properly while the run is in progress.
A turbidimeter — also called a nephelometer — shines a beam of light through the liquid sample. Particles in the liquid scatter the light in all directions. A detector positioned at 90 degrees to the light source measures how much light is scattered sideways.
The more particles, the more scattering, the higher the NTU reading.
This 90-degree measurement angle is what makes a reading "nephelometric." It is defined in international standards including ISO 7027 and US EPA Method 180.1. Both define NTU as the output of this 90-degree light-scattering measurement, though they use slightly different light sources — white light for EPA, infrared for ISO 7027 (which uses FNU — Formazin Nephelometric Units — as the unit name, though FNU and NTU are numerically equivalent for most practical purposes).
In everyday production use, the terms NTU, FNU, and FTU (Formazin Turbidity Units) are often used interchangeably. They measure the same property with the same basic technique and produce results on the same numeric scale.

Putting numbers on a scale helps make turbidity readings meaningful.
|
NTU Range |
Visual Appearance |
Example |
|
0 – 0.1 NTU |
Virtually invisible — completely clear |
Water for injection (pharmaceutical) |
|
0.1 – 1 NTU |
Clear to the naked eye, slight haze detected by instrument |
Drinking water at point of distribution |
|
1 – 5 NTU |
Slight visible haze under strong light |
Well-filtered wine or beer |
|
5 – 50 NTU |
Visible cloudiness |
Unfiltered craft beer, cloudy apple juice |
|
50 – 500 NTU |
Clearly turbid — like muddy water |
Post-fermentation wine, freshly pressed juice |
|
500 – 5,000 NTU |
Heavily turbid, opaque appearance |
Raw fruit juice, fermentation broth |
|
> 5,000 NTU |
Near-opaque |
Crude pressed oil, slurry, heavy broth |
Different industries have different NTU specifications for their finished products. These specifications drive the filtration sequence — the starting turbidity and the target turbidity determine what filtration stages are needed and at what grade.
The WHO guideline for drinking water turbidity is below 1 NTU at point of delivery, and below 0.1 NTU at the treatment plant outlet for optimal disinfection efficiency. The US EPA sets an enforceable limit of 1 NTU for filtered surface water systems, with a treatment technique target of 0.3 NTU.
Beer turbidity is commonly measured in EBC (European Brewery Convention) units, but EBC and NTU are closely related — 1 NTU ≈ 0.25 EBC for most beer types.
|
Beer Type |
Target Turbidity |
Approximate NTU Equivalent |
|
Bright filtered lager |
< 0.5 EBC |
< 2 NTU |
|
Filtered ale |
0.5 – 1 EBC |
2 – 4 NTU |
|
Hazy / unfiltered IPA |
20 – 80 EBC |
80 – 320 NTU (intentional) |
|
Turbid wheat beer |
30 – 100 EBC |
120 – 400 NTU (intentional) |
For a stainless steel multi-layer plate and frame filter for beer processing, the target output NTU determines which sheet grades are used and how many filtration stages are needed. Bright filtered beer requires polishing-grade sheets (K700 to K900). A rough beer entering the machine at 200–400 NTU leaving at under 2 NTU is a realistic and routine output from a properly configured multi-stage sheet filter.
Wine clarity is often assessed visually or by a nephelometer reading in NTU or in FAU (Formazin Attenuation Units — also numerically equivalent to NTU for most purposes).
|
Wine Clarity Level |
Approximate NTU |
|
Brilliant / star-bright |
< 1 NTU |
|
Clear |
1 – 5 NTU |
|
Slightly hazy |
5 – 20 NTU |
|
Hazy |
20 – 100 NTU |
|
Turbid |
> 100 NTU |
A finished bottled wine at retail should sit below 1–2 NTU. Post-fermentation wine before any filtration typically reads 50–500 NTU depending on how long it has settled.
Pharmaceutical specifications for injectable and ophthalmic liquids do not use NTU as the primary measure. They reference the Ph. Eur. and USP opalescence standards — reference suspensions defined in those pharmacopoeias. However, in production environments, turbidimeters calibrated in NTU are commonly used for in-process monitoring, with the understanding that the relationship between NTU readings and the pharmacopoeial opalescence standards must be established for each product.
For non-sterile oral liquids, turbidity is an appearance parameter — there is no universal NTU limit, but clarity is a specified attribute in the product's quality standard.
The incoming NTU of a liquid is one of the most reliable guides to the right sheet grade for the first filtration stage. A high starting NTU means a coarser sheet is needed. A low starting NTU means a finer sheet is appropriate from the start.
|
Incoming NTU |
Condition |
Recommended First-Stage Sheet Grade |
|
> 500 NTU |
Very turbid — crude, freshly pressed or unracked |
K100 coarse — possibly with bag pre-filter |
|
100 – 500 NTU |
Heavily turbid — post-fermentation, unfiltered |
K100 – K200 |
|
20 – 100 NTU |
Moderately turbid |
K250 – K300 |
|
5 – 20 NTU |
Slightly hazy — post-racking or pre-clarified |
K700 |
|
1 – 5 NTU |
Nearly clear — needs polishing only |
K900 |
|
< 1 NTU |
Clear — final bioburden step only |
EK / EKS |
Matching the incoming NTU to the right sheet grade from the start avoids two of the most common filtration problems — blocking fine sheets on a high-turbidity liquid, and under-filtering a low-turbidity liquid with a grade that is too coarse to achieve the target clarity.
A PLC-controlled hydraulic stainless steel sheet filter press for fine filtration can be paired with an inline turbidimeter on the outlet port. The turbidimeter monitors filtrate NTU continuously throughout the run. If filtrate NTU rises above a set threshold — indicating sheet breakthrough or bypass — the PLC triggers an alarm or redirects the filtrate back to the inlet tank for recirculation. This keeps the entire batch within specification without relying on manual sampling.
NTU is not just useful before a run — it is a practical monitoring tool throughout the filtration process.
The first fraction of filtrate from a fresh sheet set always has a higher NTU than the steady-state output. Sheet fines release at the start, and the filter cake has not yet formed inside the fiber matrix. Measuring NTU on the first fraction confirms when the filter has settled into stable operation and when it is safe to direct filtrate to the product tank rather than recirculating.
A steady or slowly improving NTU reading indicates the run is proceeding normally. A sudden rise in NTU mid-run is a warning sign — it may indicate sheet bypass, a failed plate seal, or sheets reaching saturation capacity. Catching this early prevents a batch of out-of-specification product from reaching the product tank.
NTU measured on the final fraction of the batch confirms whether the target specification has been met before the product is transferred to packaging or the next process step.
Understanding NTU values at each stage of production allows a filtration sequence to be designed with clarity. This table shows a realistic NTU profile across a multi-stage wine filtration sequence.
|
Filtration Stage |
Equipment |
Sheet Grade |
Inlet NTU |
Target Outlet NTU |
|
First clarification |
Sheet filter press |
K100 |
200 – 400 |
20 – 50 |
|
Second clarification |
Sheet filter press |
K250 |
20 – 50 |
5 – 15 |
|
Fine polishing |
Sheet filter press |
K700 |
5 – 15 |
1 – 3 |
|
Pre-bottling polish |
Sheet filter press |
K900 |
1 – 3 |
< 1 |
Each stage takes the output of the previous one and reduces turbidity further. The sheet grade at each stage is matched to the NTU of the liquid arriving at that stage — not the raw incoming liquid and not the final target.
For process development or when establishing NTU profiles for a new product or new season's crop, a lab-scale mini plate and sheet filter allows each stage to be tested at small volume. Measuring inlet and outlet NTU at each stage on a lab machine gives the full NTU reduction profile before the production sequence is committed to a full-scale machine configuration.
The full range of stainless steel plate and sheet filter presses covers every production scale from small craft operations to high-volume commercial lines, with sheet grade flexibility from K100 coarse through to EK sterilising grade — the full span of the NTU reduction range needed for food, beverage, and pharmaceutical liquid production.